中文   |   English   Hance Technologies Limited           Tel:  021-57863847
ammonia ETC for PV modules

ammonia ETC for PV modules



Ammonia Test Chamber for PV modules
 
Main Specifications
 
Temperature control range: 15 ℃ ~ 70 ℃,
Temperature fluctuation: ≤ ± 0.5 ℃,
Temperature deviation: ≤ ± 1.0 ℃,
Temperature uniformity: ≤ ± 2 ℃;
Humidity control range: 40 ~ 100% RH adjustable,
Humidity deviation: ≤ ± 2% RH,
Humidity uniformity: ≤ ± 3% RH.
NH3 concentration: 0-10000ppm adjustable,
Ammonia gas concentration test resolution: ± 1ppm,
Test accuracy ± 2%.
 
The test chamber shall meet the technical requirements of the latest version of the test standards such as IEC61701, IEC60068-2-52, GBT2423.17, GBT2423.18, IEC 62716 (not yet released, existing draft version is 82/722/CDV). The device can completely support all testing processes required by IEC61701 and IEC draft standard 82/722/CDV. That is to say, the harsh test grade 1-6 spray, damp-heat storage, standard atmospheric environment and sample cleaning stage are all completed in the NaCl corrosion chamber; ammonia, damp and heat storage, spray cleaning, drying can all be done in the NH3 corrosion chamber.
 

Links:   

Tel:021-57863847     E-mail: sales@hance-tech.com

Add:Yuxiu Rd 136, Songjiang, 201699 Shanghai

Hance Technologies Ltd, based in Shanghai, with an area of more than 3,500 sqm of proprietary facility, show room and office floor on the premises covering an area about 7000 sqm, with representative offices in major Chinese cities such as Hongkong, Canton, Peking, Chungking and Wuhan etc., is a comp...